REOCOM 4 channels
REOCOM 4 channels

Digital 4-channel rheograph

  • Rheoencephalography (REG)
  • Rheovasography (RVG)
  • Central hemodynamics
Request a quote
REOCOM 8 channels
REOCOM 8 channels

Digital 8-channel rheograph

  • Rheoencephalography (REG), rheovasography (RVG), central hemodynamics
  • Occlusion rheography and systolic limb blood flow
  • Endothelial function assessment
  • Blood pressure and ankle-brachial index (ABI)
  • Rheo-ophthalmography
Request a quote

Key specifications

REO channels
4 / 8
ECG channel
1
Method
tetrapolar
Base impedance
1–1000 Ω
AC component
0.002–5 Ω
Current
1 mA, 16 / 32 / 64 / 128 kHz
Phono-ECG
by kit
Interface
USB

Public procurement codes (Prozorro)

Ready-to-use wording for Ukrainian tender documents, specifications and contracts. We can send technical requirements on request.

Комплекс реографічний комп'ютерний для дослідження центральної та периферичної гемодинаміки

ДК 021:2015
33120000-7 Системи реєстрації медичної інформації та дослідне обладнання
НК 024:2023
35176 Кардіоаналізатор багатофункціональний
НК 031:2024
Z12050101 Прилади для моніторингу фізіологічного стресу / функціональних проб

Software capabilities

Software REOCOM

  • Patient file: multiple databases on one or several media; network use (record and review on different PCs); search, sort, statistics; shared base with EEG / SPIRO; study import / export
  • Study-type and test-protocol editor: number of tests, lead schemes, sensitivity / speed, duration, stop conditions, report template and norm table
  • Lead-scheme editor and open norm modules; create and edit age–sex norms and report templates
  • Recording: volume rheogram, base impedance and differential curve; sync ECG and phono channels; 1–8 rheo channels; central hemodynamics synced with other vascular beds
  • LF / HF filtering (order and cut-offs); mains filter; automatic isoelectric correction (respiratory wave removal)
  • Rheogram split into LF and HF components; joint amplitude–phase analysis of tone, peripheral resistance and flow
  • Coherent averaging by correlation coefficient — whole test, fixed interval or sliding window
  • Automatic landmark detection with manual marker correction; Markers window
  • 80+ automatically calculated parameters; selectable indicators for the result table and auto description; comparison with norms
  • Auto description and conclusion editor; full / short / model («m») templates; facility and physician details in the report
  • Spectral analysis of rheo signals (Spectrum window)
  • Export of raw data and calculated parameters (Word, Excel); network transfer
  • Integration with medical information systems (HIS / MIS)
  • Our own solution for telemedicine consultations, storage and data exchange — CLORM
  • Interface, reports and printouts in Ukrainian, English and Russian

Pro · 8 channels

  • Occlusion rheography and systolic limb blood-flow volume
  • Automatic endothelial function assessment (occlusion / nitroglycerin tests)
  • Blood pressure and ankle-brachial index (ABI), including vasodilatation tests
  • Pulse-wave velocity (PWV)
  • Rheo-ophthalmography
  • Organ rheography (lungs, liver, etc.)

Request a quote

Documents and guides

Quick guides

Minimizing skin–electrode impedance

ECG · EEG · Holter · rheography · quick guide

≤ 5 kΩEEG: impedance of each electrode (check in the software)
5 stepsskin preparation before applying electrodes
1 usedisposable electrode: never reuse
No prepartifacts, baseline drift, mains noise

1Why impedance must be low

No preparation ElectrodeGelStratum corneumEpidermisDermisFatRgelRscRsub Z high After preparation ElectrodeGelStratum corneumEpidermisDermisFatRgelRscRsub Z low

Most of the impedance comes from the stratum corneum (dead cells, oil, dust). The higher it is, the more mains noise and baseline drift, especially on microvolt-level signals (EEG).

2Skin preparation: 5 steps

  1. 1Siteaway from muscle, folds, scars and wounds
  2. 2Hairshave or part it
  3. 3Cleanalcohol, let it dry completely
  4. 4Abraderub lightly until slightly pink
  5. 5Electrodefresh, with gel (RVG: moistened); press firmly, check impedance

3By examination type

ECG

  • Ag/AgCl electrodes, not dried out, within expiry date
  • Cable without strain; patient relaxed and not talking

EEG

  • Each electrode ≤ 5 kΩ and roughly equal
  • Part the hair, gel/paste, cap fitted snugly
  • Check impedances in the software before recording

Holter

  • Shave, always degrease and lightly abrade the skin
  • Secure with tape, leave a cable loop (strain relief)
  • For long recordings replace electrodes per manufacturer’s guidance

Rheography (REG, RVG)

  • REG: gel electrodes over the whole contact area
  • RVG: moisten the electrode/pad, do not let it dry out
  • Firm fixing (strap/tape), no folds or air

✔ Pre-recording checklist

Skin cleaned and fully dry
Hair removed
Electrodes fresh, with gel (RVG: moistened), in date
Firm contact, no folds or air
Impedance checked in software (EEG ≤ 5 kΩ)
Cables without strain, away from mains cords
Do not “fix” high impedance with gain or filters — re-apply the electrode.Never reuse disposable electrodes, mix new and old ones, or apply to irritated or damaged skin.

Electrode contact · guide · PDF 262 KB

Electrode placement: REG, RVG and central haemodynamics

RheoCom · REG · RVG · central haemodynamics · quick guide

ECG + Nmandatory; the R wave must be positive
≥ 1 cmbetween current (I) and nearest voltage (U) electrode
Symmetricelectrodes on both sides
Measurethe electrode distance and enter it after recording

1General rules

  • Current (I) and voltage (U) electrodes are separate; keep them at least 1 cm apart.
  • ECG electrodes and N — mandatory. The R wave is positive and clearly visible.
  • Symmetrically on both sides.
  • Firmly, without squeezing vessels: elastic band, even tension.
  • Posture: supine (head slightly raised) or sitting — note it in the protocol. When sitting, a neck rest is preferable to relax the neck muscles.
  • No artefacts: relax, eyes closed, no talking, no swallowing.

2REG — rheoencephalography

REG — rheoencephalography

top and side view

  • As for EEG: clean and degrease the skin; gel only within the contact surface.
  • Occiput: part the hair, degrease especially well.
  • I — mid-forehead and mid-occiput.
  • U1, U3 — on the right side, U2, U4 — on the left (symmetrically).
  • FM (fronto-mastoid) — carotid arteries; OM (occipito-mastoid) — vertebral arteries.
  • Measure the distance between the U electrodes — and always enter it.

3RVG — limb rheovasography

RVG — limb rheovasographyRVG — limb rheovasography

arms and legs

  • Moisten the tape electrode — the surface touching the body.
  • I — distal, U — proximal.
  • U1 (U3) — on the right side, U2 (U4) — on the left. U3 and U4 — only when recording four channels at once.
  • Measure the distance between the U electrodes — and always enter it.

4Central haemodynamics

Central haemodynamics

central haemodynamics

  • Thoracic rheography (Kubicek): two tapes on the neck near the clavicles, two around the lower chest.
  • Integral (Tishchenko): tapes on the distal forearms and calves.
  • Pushkar variant (figure): U — neck and chest, I — head and leg.
  • Record ECG synchronously.
  • Select SV (Kubicek) in the parameters.
  • Measure the distance between the U electrodes — and always enter it.

5Workflow in the program

1Fill in a new patient card or select one from the card index.

2Record a new study by selecting its type.

Record a new study by selecting its type.

3After you stop recording, the “Settings” window opens — enter:

undefined
Height, weight — taken from the patient card. If the real values differ (growing children), enter the actual ones.
Neck and chest circumference — for central haemodynamics.
BP — measure it if possible, or enter the value the patient knows.
Electrode distance — the measured value, mandatory.
undefined
  • Height, weight — taken from the patient card. If the real values differ (growing children), enter the actual ones.
  • Neck and chest circumference — for central haemodynamics.
  • BP — measure it if possible, or enter the value the patient knows.
  • Electrode distance — the measured value, mandatory.
Always enter the measured electrode distance.Without it amplitude indices and velocities are calculated wrongly.

Electrode placement · guide · PDF 389 KB

Rheogram analysis: standard approach and LF/HF decomposition

RheoCom · user guide · 3 pages

Page 1: standard rheogram analysis — markers a, b, c, i, d, parameters ΔR, Rz, td, α1, α2, MCi, MCd
Page 2: where interpretation diverges — ambiguous marker c, amplitude dependence on distance and cross-section
Page 3: LF and HF decomposition, pulse wave phase PWP, comparison with the standard analysis

Rheogram analysis · guide · PDF 388 KB

More on the LF and HF method — read here Article in the Publications section

General earthing requirements for functional diagnostics systems

Earthing of functional diagnostics devices

Standards: IEC 60601-1, IEC 62353, local wiring rules

≤ 2 Ωworking earth (FE) loop resistance
≤ 0.1 Ωcontact resistance: device clamp — bar
≥ 15 mfrom other earth electrodes (lightning protection, lift)
TN-S / ITTN-C (PEN) prohibited

1Mains: TN-S or IT only

✔ TN-S (correct) L N PE PC PE and N are separate ✖ TN-C (prohibited) L PEN PC PE merged with N — noise!

The protective conductor is the third (earth) pin of the socket. Never fit a PE→N jumper in the room. IT — via a medical isolating transformer.

2“Star” layout, not “daisy chain”

✔ Star BAR PCShield* * if present ✖ Daisy chain bar PCShield* in series → ground loops

Each device goes to the bar with its own conductor. Loops pick up mains noise into the microvolt-level signal.

3Equipotential bonding in the room

EQUIPOTENTIAL BONDING BAR Device PCRoom shield* * if present to the separate FE loop ≤ 2 Ω Braided copper strap between earthing points fewer artifacts in the record Contact ≤ 0.1 Ω clamp → bar: tighten, clean, no paint/oxide

4Earthing loop requirements

  • A separate working (functional) earth, FE, independent of the hospital protective earth (protective earth norm: up to 4 Ω).
  • Distance to other earthing systems — at least 15 m.
  • Connect only through the third (earth) conductor of the socket.

5Testing and operation

  • Measure contact resistances and insulation before first use.
  • Then periodically per clinic policy (usually once a year).
  • Record the results in a log.

✔ Pre-start checklist

Mains is TN-S or IT, no PEN in the room
Device PC chassis — on the bar
Room shield (if present) — on the bar
“Star” connection, no daisy chains
FE loop ≤ 2 Ω (measured)
Clamp contact resistances ≤ 0.1 Ω
Do not switch the device on if the earthing has not been measured or the socket has no earth conductor.Prohibited: PEN conductor, PE→N jumper in the socket, daisy-chained connection of devices.

Earthing · guide · PDF 93 KB

Frequently asked questions

What is REOCOM for?

REOCOM is a computer rheograph for regional blood-flow and central hemodynamics assessment in functional diagnostics, cardiology, neurology and vascular clinics. Clinically it is used for rheoencephalography (REG), rheovasography (RVG) and central hemodynamics; Pro · 8 channels adds occlusion tests, endothelial function, blood pressure and ABI, organ rheography and rheo-ophthalmography. Data integrate with the facility HIS.

Does it integrate with HIS?

Yes. Results go to the facility’s medical information systems (HIS / MIS). Separately, our own solution for telemedicine consultations, storage and data exchange — CLORM — is available.

Price, quote, warranty and service: what is included and how to order?

The public price list is in Resources, PDF: price list.

For a quote, use “Request a quote” or message Telegram / Viber / WhatsApp +38 095 047-4000. The same applies to a custom configuration.

The price includes delivery, commissioning and staff training.

24-month warranty, with service support throughout the entire operating life of the equipment.