4- and 8-channel rheograph
REOCOM — 4- or 8-channel rheograph
REG, RVG and central hemodynamics; Pro · 8 channels adds occlusion, endothelium, BP, ankle-brachial index (ABI) and other vascular tests.
Digital 4-channel rheograph
- Rheoencephalography (REG)
- Rheovasography (RVG)
- Central hemodynamics
Digital 8-channel rheograph
- Rheoencephalography (REG), rheovasography (RVG), central hemodynamics
- Occlusion rheography and systolic limb blood flow
- Endothelial function assessment
- Blood pressure and ankle-brachial index (ABI)
- Rheo-ophthalmography
Key specifications
- REO channels
- 4 / 8
- ECG channel
- 1
- Method
- tetrapolar
- Base impedance
- 1–1000 Ω
- AC component
- 0.002–5 Ω
- Current
- 1 mA, 16 / 32 / 64 / 128 kHz
- Phono-ECG
- by kit
- Interface
- USB
Public procurement codes (Prozorro)
Ready-to-use wording for Ukrainian tender documents, specifications and contracts. We can send technical requirements on request.
Комплекс реографічний комп'ютерний для дослідження центральної та периферичної гемодинаміки
- ДК 021:2015
33120000-7Системи реєстрації медичної інформації та дослідне обладнання- НК 024:2023
35176Кардіоаналізатор багатофункціональний- НК 031:2024
Z12050101Прилади для моніторингу фізіологічного стресу / функціональних проб
Software capabilities

Software REOCOM
- Patient file: multiple databases on one or several media; network use (record and review on different PCs); search, sort, statistics; shared base with EEG / SPIRO; study import / export
- Study-type and test-protocol editor: number of tests, lead schemes, sensitivity / speed, duration, stop conditions, report template and norm table
- Lead-scheme editor and open norm modules; create and edit age–sex norms and report templates
- Recording: volume rheogram, base impedance and differential curve; sync ECG and phono channels; 1–8 rheo channels; central hemodynamics synced with other vascular beds
- LF / HF filtering (order and cut-offs); mains filter; automatic isoelectric correction (respiratory wave removal)
- Rheogram split into LF and HF components; joint amplitude–phase analysis of tone, peripheral resistance and flow
- Coherent averaging by correlation coefficient — whole test, fixed interval or sliding window
- Automatic landmark detection with manual marker correction; Markers window
- 80+ automatically calculated parameters; selectable indicators for the result table and auto description; comparison with norms
- Auto description and conclusion editor; full / short / model («m») templates; facility and physician details in the report
- Spectral analysis of rheo signals (Spectrum window)
- Export of raw data and calculated parameters (Word, Excel); network transfer
- Integration with medical information systems (HIS / MIS)
- Our own solution for telemedicine consultations, storage and data exchange — CLORM
- Interface, reports and printouts in Ukrainian, English and Russian
Pro · 8 channels
- Occlusion rheography and systolic limb blood-flow volume
- Automatic endothelial function assessment (occlusion / nitroglycerin tests)
- Blood pressure and ankle-brachial index (ABI), including vasodilatation tests
- Pulse-wave velocity (PWV)
- Rheo-ophthalmography
- Organ rheography (lungs, liver, etc.)
Documents and guides
Downloads
Quick guides
Minimizing skin–electrode impedance
ECG · EEG · Holter · rheography · quick guide
1Why impedance must be low
Most of the impedance comes from the stratum corneum (dead cells, oil, dust). The higher it is, the more mains noise and baseline drift, especially on microvolt-level signals (EEG).
2Skin preparation: 5 steps
- 1Siteaway from muscle, folds, scars and wounds
- 2Hairshave or part it
- 3Cleanalcohol, let it dry completely
- 4Abraderub lightly until slightly pink
- 5Electrodefresh, with gel (RVG: moistened); press firmly, check impedance
3By examination type
ECG
- Ag/AgCl electrodes, not dried out, within expiry date
- Cable without strain; patient relaxed and not talking
EEG
- Each electrode ≤ 5 kΩ and roughly equal
- Part the hair, gel/paste, cap fitted snugly
- Check impedances in the software before recording
Holter
- Shave, always degrease and lightly abrade the skin
- Secure with tape, leave a cable loop (strain relief)
- For long recordings replace electrodes per manufacturer’s guidance
Rheography (REG, RVG)
- REG: gel electrodes over the whole contact area
- RVG: moisten the electrode/pad, do not let it dry out
- Firm fixing (strap/tape), no folds or air
✔ Pre-recording checklist
Electrode placement: REG, RVG and central haemodynamics
RheoCom · REG · RVG · central haemodynamics · quick guide
1General rules
- Current (I) and voltage (U) electrodes are separate; keep them at least 1 cm apart.
- ECG electrodes and N — mandatory. The R wave is positive and clearly visible.
- Symmetrically on both sides.
- Firmly, without squeezing vessels: elastic band, even tension.
- Posture: supine (head slightly raised) or sitting — note it in the protocol. When sitting, a neck rest is preferable to relax the neck muscles.
- No artefacts: relax, eyes closed, no talking, no swallowing.
2REG — rheoencephalography

top and side view
- As for EEG: clean and degrease the skin; gel only within the contact surface.
- Occiput: part the hair, degrease especially well.
- I — mid-forehead and mid-occiput.
- U1, U3 — on the right side, U2, U4 — on the left (symmetrically).
- FM (fronto-mastoid) — carotid arteries; OM (occipito-mastoid) — vertebral arteries.
- Measure the distance between the U electrodes — and always enter it.
3RVG — limb rheovasography


arms and legs
- Moisten the tape electrode — the surface touching the body.
- I — distal, U — proximal.
- U1 (U3) — on the right side, U2 (U4) — on the left. U3 and U4 — only when recording four channels at once.
- Measure the distance between the U electrodes — and always enter it.
4Central haemodynamics

central haemodynamics
- Thoracic rheography (Kubicek): two tapes on the neck near the clavicles, two around the lower chest.
- Integral (Tishchenko): tapes on the distal forearms and calves.
- Pushkar variant (figure): U — neck and chest, I — head and leg.
- Record ECG synchronously.
- Select SV (Kubicek) in the parameters.
- Measure the distance between the U electrodes — and always enter it.
5Workflow in the program
1Fill in a new patient card or select one from the card index.
2Record a new study by selecting its type.

3After you stop recording, the “Settings” window opens — enter:

- Height, weight — taken from the patient card. If the real values differ (growing children), enter the actual ones.
- Neck and chest circumference — for central haemodynamics.
- BP — measure it if possible, or enter the value the patient knows.
- Electrode distance — the measured value, mandatory.
Rheogram analysis: standard approach and LF/HF decomposition
RheoCom · user guide · 3 pages



Rheogram analysis · guide · PDF 388 KB
More on the LF and HF method — read here Article in the Publications section
General earthing requirements for functional diagnostics systems
Earthing of functional diagnostics devices
Standards: IEC 60601-1, IEC 62353, local wiring rules
1Mains: TN-S or IT only
The protective conductor is the third (earth) pin of the socket. Never fit a PE→N jumper in the room. IT — via a medical isolating transformer.
2“Star” layout, not “daisy chain”
Each device goes to the bar with its own conductor. Loops pick up mains noise into the microvolt-level signal.
3Equipotential bonding in the room
4Earthing loop requirements
- A separate working (functional) earth, FE, independent of the hospital protective earth (protective earth norm: up to 4 Ω).
- Distance to other earthing systems — at least 15 m.
- Connect only through the third (earth) conductor of the socket.
5Testing and operation
- Measure contact resistances and insulation before first use.
- Then periodically per clinic policy (usually once a year).
- Record the results in a log.