NEUROCOM 16 / 19 EEG
NEUROCOM — 16 / 19-channel electroencephalograph
Digital EEG in NEUROCOM 16 and NEUROCOM 19 configurations: routine recording, functional tests and evoked potentials.
Digital 16-channel EEG
- Routine EEG
- Hyperventilation
- Photic stimulation
- IFCN protocol
- Auditory and visual EP
- EEG caps via a separate connector
Digital 19-channel EEG
- Routine EEG
- Hyperventilation
- Photic stimulation
- IFCN protocol
- Auditory and visual EP
- EEG caps via a separate connector
- Standard 10–20 system
- 4 HF channels
- ECG channel
Key specifications
- Sampling rate
- 4000 / 500 Hz
- EEG range
- 1–12,000 µV
- Resolution
- 24 bit
- Bandwidth
- 0.5–50 Hz
- Internal noise
- < 1 µV
- Input impedance
- diff. > 10 MΩ · common-mode > 100 MΩ
- CMRR
- >120 dB
- Interface
- USB
- Routine EEG mode
- IFCN: continuous recording, test markers
- Photic stimulation (IPS)
- 1–30 Hz, 13 frequencies in the IFCN protocol
- Additional channels (NEUROCOM 19)
- EMG + ECG in sync with EEG
- Video (Pro · NEUROCOM 19)
- Recorded in sync with EEG and markers
Public procurement codes (Prozorro)
Ready-to-use wording for Ukrainian tender documents, specifications and contracts. We can send technical requirements on request.
Комплекс електроенцефалографічний комп'ютерний
- ДК 021:2015
33120000-7Системи реєстрації медичної інформації та дослідне обладнання- НК 024:2023
11467Електроенцефалограф- НК 031:2024
Z12100302Електроенцефалографи
Five advantages of the IFCN mode
-
Continuous EEG recording per the IFCN protocol The whole study is one gap-free file; tests (eyes open / closed, photic stimulation) are flagged with event markers while recording.
-
Protocol-driven photic stimulation Rhythmic photic stimulation at 1–30 Hz: 13 frequencies of 5 s each, triggered at the right moment of the recording; stimulus-locked artifacts are removed.
-
Additional EMG channels Myogram recorded in sync with the EEG — easily tell muscle artifacts apart and assess muscle activity (NEUROCOM 19).
-
ECG channel on the trace ECG on the same time axis as the EEG helps tell cardiac artifacts apart and monitor rhythm during recording (NEUROCOM 19).
-
Synchronized video monitoring Patient video is recorded with the EEG and event markers — behaviour during episodes is visible next to the trace; the file can be deleted after review (Pro · NEUROCOM 19).
Software capabilities

Software NEUROCOM
- Patient file: multiple databases on one or several media; network use (record and review on different PCs); search, sort, statistics; shared base with REO / SPIRO; study import / export
- Study-type and test-protocol editor: number of tests, lead schemes, stimulation, sensitivity / speed, duration, report template and norm table
- Montage and lead-scheme editor (built-in and user); spatial filtering of EEG leads
- IFCN mode (European protocol): continuous EEG recording in one file, tests flagged with event markers; rhythmic photic stimulation at 1–30 Hz per the protocol; additional EMG and ECG channels and synchronized video (NEUROCOM 19 / Pro)
- Protocol EEG recording: resting, activation reaction, rhythmic photostimulation, hyperventilation; automatic test advance; event markers
- Filtering of all or selected leads (LP / HP / band-pass, custom cut-offs and order); zero-phase symmetric filters; adaptive mains comb
- ICA / BSS (Independent Component Analysis / Blind Source Separation): blind source separation; ICA-component extraction
- Artifact rejection without shortening the record (electrode and physiological) by subtracting artifact components; for EP — removal of stimulus-locked artifacts
- EEG-phenomenon markup by class with amplitude, duration, period and localization; automatic rhythm parameters and auto description with report editor
- Spectral analysis of EEG / EP / ICA (resolution and window choice); correlation and coherence analysis
- Topographic mapping of EEG, EP and ICA: power, frequency, potential, dipole and symmetry maps; map families; driven-rhythm maps and harmonics
- Spatial localization of individual phenomena / components; 3D tomography; comparison with the patient’s MRI
- Flash VEP; separate odd/even stimulus averaging; accumulation-quality control; EP re-averaging with ICA
- Integration with medical information systems (HIS / MIS)
- Our own solution for telemedicine consultations, storage and data exchange — CLORM
- Interface, reports and printouts in Ukrainian, English and Russian
Pro software — NEUROCOM 19 only
- Patient video monitoring: image recorded in sync with EEG and event markers for later behaviour review; the file can be deleted after processing
- Checkerboard and arbitrary on-display image stimulation; fibre-optic sync with a second monitor
- Long-latency auditory EP (DSVP), cognitive P300, short-latency BAEP; separate averaging and quality control
- Biofeedback: real-time training on amplitudes of selected EEG rhythms
- Brain-death confirmation under a dedicated study protocol
Documents and guides
Quick guides
Minimizing skin–electrode impedance
ECG · EEG · Holter · rheography · quick guide
1Why impedance must be low
Most of the impedance comes from the stratum corneum (dead cells, oil, dust). The higher it is, the more mains noise and baseline drift, especially on microvolt-level signals (EEG).
2Skin preparation: 5 steps
- 1Siteaway from muscle, folds, scars and wounds
- 2Hairshave or part it
- 3Cleanalcohol, let it dry completely
- 4Abraderub lightly until slightly pink
- 5Electrodefresh, with gel (RVG: moistened); press firmly, check impedance
3By examination type
ECG
- Ag/AgCl electrodes, not dried out, within expiry date
- Cable without strain; patient relaxed and not talking
EEG
- Each electrode ≤ 5 kΩ and roughly equal
- Part the hair, gel/paste, cap fitted snugly
- Check impedances in the software before recording
Holter
- Shave, always degrease and lightly abrade the skin
- Secure with tape, leave a cable loop (strain relief)
- For long recordings replace electrodes per manufacturer’s guidance
Rheography (REG, RVG)
- REG: gel electrodes over the whole contact area
- RVG: moisten the electrode/pad, do not let it dry out
- Firm fixing (strap/tape), no folds or air
✔ Pre-recording checklist
General earthing requirements for functional diagnostics systems
Earthing of functional diagnostics devices
Standards: IEC 60601-1, IEC 62353, local wiring rules
1Mains: TN-S or IT only
The protective conductor is the third (earth) pin of the socket. Never fit a PE→N jumper in the room. IT — via a medical isolating transformer.
2“Star” layout, not “daisy chain”
Each device goes to the bar with its own conductor. Loops pick up mains noise into the microvolt-level signal.
3Equipotential bonding in the room
4Earthing loop requirements
- A separate working (functional) earth, FE, independent of the hospital protective earth (protective earth norm: up to 4 Ω).
- Distance to other earthing systems — at least 15 m.
- Connect only through the third (earth) conductor of the socket.
5Testing and operation
- Measure contact resistances and insulation before first use.
- Then periodically per clinic policy (usually once a year).
- Record the results in a log.