Holter ECG and combined ECG+BP system
CARDIOSENS — Holter ECG and combined ECG+BP
Two options: a digital 3-channel Holter ECG and a separate combined ECG+BP system. A 4G telemetry option is available.
Digital 3-channel Holter ECG
- 24–72 hours of ECG
- Miniature size, 60 g
- Only 4 wires for 3 leads
- Built-in battery and built-in 32 GB memory
Combined ECG + BP system
- 24–72 hours of ECG
- 24 hours of BP readings
- Miniature size, up to 200 g
- Only 4 wires for 3 leads
- Built-in battery and built-in 32 GB memory
Key specifications
- ECG recording
- 24–72 hours, 3 channels
- Input signal range
- ± 8 mV
- Bandwidth
- 0.5–75 Hz
- CMRR
- 110 dB
- Internal noise
- < 5 µV
- Weight
- ≈ 60 g (ECG) · ≈ 200 g (ECG+BP)
- BP (BP version)
- up to ~250 readings; adult and pediatric modes
- BP range
- 10–280 mmHg (adults) · 5–155 mmHg (children)
- Memory
- built-in, 32 GB
- Interfaces
- USB-C, Bluetooth
Public procurement codes (Prozorro)
Ready-to-use wording for Ukrainian tender documents, specifications and contracts. We can send technical requirements on request.
Комплекс моніторингу електрокардіосигналів (амбулаторний реєстратор за методом Холтера)
- ДК 021:2015
33120000-7Системи реєстрації медичної інформації та дослідне обладнання- НК 024:2023
35162Реєстратор амбулаторний для тривалого електрокардіографічного моніторингу- НК 031:2024
Z12050402Реєстратори ЕКГ за холтером
Комплекс комбінованого амбулаторного моніторингу електрокардіосигналів та артеріального тиску з функцією автоматичного вимірювання артеріального тиску
- ДК 021:2015
33120000-7Системи реєстрації медичної інформації та дослідне обладнання- НК 024:2023
35162Реєстратор амбулаторний для тривалого електрокардіографічного моніторингу- НК 031:2024
Z12050402Реєстратори ЕКГ за холтером
Software capabilities

Software CARDIOSENS
- Patient file: multiple databases and folders, monitoring diary, network use
- Recorder programming and control ECG (USB-C / Bluetooth), record download to PC over USB
- Artifact filtering, QRS detection and markup
- Morphological classification (normal, SVE, VE, aberrant, unclassified, artifacts; with a pacemaker — spontaneous / paced / fusion), classifier training and manual cluster editing
- P waves, P–Q intervals, atrial activity; scatterplot; electronic magnifier
- Automatic rhythm and conduction analysis: counts per 24 h, hour, day and night; episode descriptions; dictionary and manual edits
- Rhythm: tachycardia, bradycardia, pauses; atrial and ventricular fibrillation / flutter
- SVE and VE: single, interpolated, couplets, polymorphic, bi-/tri-/quadrigeminy, runs; SVT and VT; R-on-T
- Conduction disorders and AV block; high-grade ectopy (Lown 4b, 5); automatic pacemaker-pulse detection
- ST-T with decision-rule and episode-boundary correction; QRS morphology; Q–T intervals
- Heart rate variability (HRV) in time and frequency domains (ESC working-group norms, PARS — regulatory systems activity index); AC/DC via PRSA; heart-rate turbulence (HRT)
- EASI-12 reconstruction of 12 leads from the 3-channel recording
- Auto report with ECG strips, conclusion-form editor, print
- Export of records and tables (network, e-mail, Excel, MATLAB)
- HIS integration
- Integration with our own shared-access and data-storage system CLORM
- Interface, reports and printouts in Ukrainian, English and Russian
ECG+BP version
- BP per NBREP: SBP / DBP / mean / pulse pressure and HR trends; cuff oscillation and pulse-wave review
- Distribution histograms, diagrams, correlation, cosinor analysis
- Pressure-load indices (measurement, time, area); BP variability (VAR1–3)
- Nocturnal dip (SNS), day / night / custom intervals; morning-surge magnitude and rate
- Calibration BP readings and offsets; adult and pediatric protocols
- Summary tables and a separate BP report (form editor, fragment and full-report print)
Documents and guides
Downloads
Quick guides
Minimizing skin–electrode impedance
ECG · EEG · Holter · rheography · quick guide
1Why impedance must be low
Most of the impedance comes from the stratum corneum (dead cells, oil, dust). The higher it is, the more mains noise and baseline drift, especially on microvolt-level signals (EEG).
2Skin preparation: 5 steps
- 1Siteaway from muscle, folds, scars and wounds
- 2Hairshave or part it
- 3Cleanalcohol, let it dry completely
- 4Abraderub lightly until slightly pink
- 5Electrodefresh, with gel (RVG: moistened); press firmly, check impedance
3By examination type
ECG
- Ag/AgCl electrodes, not dried out, within expiry date
- Cable without strain; patient relaxed and not talking
EEG
- Each electrode ≤ 5 kΩ and roughly equal
- Part the hair, gel/paste, cap fitted snugly
- Check impedances in the software before recording
Holter
- Shave, always degrease and lightly abrade the skin
- Secure with tape, leave a cable loop (strain relief)
- For long recordings replace electrodes per manufacturer’s guidance
Rheography (REG, RVG)
- REG: gel electrodes over the whole contact area
- RVG: moisten the electrode/pad, do not let it dry out
- Firm fixing (strap/tape), no folds or air
✔ Pre-recording checklist
CARDIOSENS: electrodes and recording start
Holter ECG · ECG + BP · quick guide
1Electrode placement: 4 electrodes
- White — left subclavicular area (−)
- Red — right, 4th intercostal space, position V1 (1+)
- Yellow — position V4 (2+)
- Green — position V5 (3+)
The recorded signals roughly match the leads: 3 channels: 1 — V1, 2 — aVF, 3 — V5 (EASI-12 optional).
2Recorder and lead cable

- Recorder: in a pouch on the belt or on a shoulder strap, not in a pocket
- Leave a cable loop at each electrode (strain relief) and tape the cable down
- Keep cables slack: big loops, no strain or twisting
- No synthetic vests or T-shirts (static); prefer natural fabrics
3Starting a recording: 5 steps
- 1Patientselect or create a record in the database
- 2Recorder“Recorder preparation”: mode, scheme, time; write
- 3Electrodesprepare the skin, apply per the scheme, connect the cable
- 4Control ECG3 clean channels; otherwise re-apply the electrode
- 5Startstart monitoring, note the time, hand over the diary
4BP version: the cuff
- Cuff size by upper-arm circumference; adult / child protocol; control reading and correction in the software
- Place on the upper arm, lower edge 2–3 cm above the elbow crease; snug, one finger fits under it
- Cuff on the left arm; run the tube across the chest, behind the neck and down the shoulder to the cuff — slack, no strain
- During a measurement: arm relaxed and still, at heart level
5What to tell the patient
- With pain, breathlessness or palpitations press the patient button on the recorder
- Keep a diary: sleep, meals, activity, medication, symptoms — with exact times
- Do not remove or wet it; no electric blankets
- If an electrode comes off, press it back or replace it and note the time in the diary
✔ Checklist before the patient leaves
General earthing requirements for functional diagnostics systems
Earthing of functional diagnostics devices
Standards: IEC 60601-1, IEC 62353, local wiring rules
1Mains: TN-S or IT only
The protective conductor is the third (earth) pin of the socket. Never fit a PE→N jumper in the room. IT — via a medical isolating transformer.
2“Star” layout, not “daisy chain”
Each device goes to the bar with its own conductor. Loops pick up mains noise into the microvolt-level signal.
3Equipotential bonding in the room
4Earthing loop requirements
- A separate working (functional) earth, FE, independent of the hospital protective earth (protective earth norm: up to 4 Ω).
- Distance to other earthing systems — at least 15 m.
- Connect only through the third (earth) conductor of the socket.
5Testing and operation
- Measure contact resistances and insulation before first use.
- Then periodically per clinic policy (usually once a year).
- Record the results in a log.