Digital 12-channel ECG
CARDIOLAB — 12-channel electrocardiograph
A computer electrocardiograph for clinics, hospitals and sports medicine. Integrates with the facility HIS.

Digital 12-channel ECG
- Resting ECG with auto-interpretation
- Vector ECG analysis
- Heart rate variability (HRV) and ANS assessment (Baevsky / PARS — regulatory systems activity index)
- Cycle ergometry and treadmill tests
- Phonocardiography
- Athlete assessment after Dushanin
Key specifications
- Recording range
- ± 330 mV
- Frequency range
- 0.05–200 Hz
- Resolution
- 24 bit
- Internal noise
- < 5 µV
- Interfaces
- USB, Bluetooth
- Other
- Pacemaker detection
Public procurement codes (Prozorro)
Ready-to-use wording for Ukrainian tender documents, specifications and contracts. We can send technical requirements on request.
Комплекс електрокардіографічний комп'ютерний багатоканальний
- ДК 021:2015
33120000-7Системи реєстрації медичної інформації та дослідне обладнання- НК 024:2023
16231Професійний багатоканальний електрокардіограф- НК 031:2024
Z12050302Електрокардіографи розширеної діагностики
Software capabilities

Software CARDIOLAB
- Patient file: archives, folders, cards, search, statistics
- Standard ECG recording (10 s – 2 min) in 12 standard leads and in Frank, Neb, Slopak and other lead systems
- Automatic PQRST markup, parameter table, manual marker correction
- Automatic syndromic interpretation; conclusion editor and custom forms
- ST and QT analysis
- Vector ECG: F / Sl / H projections and 3D; Frank leads or transform from 12 leads
- ECG recording and analysis for patients with an implanted pacemaker
- HRV: time-domain, spectral and geometric analysis (ESC WG 1996 and Baevsky)
- Autonomic tone assessment and composite PARS score
- ANS tests: orthostatic, deep breathing, Valsalva, exercise; automatic ANS conclusion
- Phonocardiography: heart sounds and murmurs, electrical / mechanical systole; time–frequency analysis; Maas–Weber band playback; sound library
- Ergometry: standard and user protocols (cycle / treadmill); load control
- During exercise: averaged QRS, ST level and slope, HR, alarm parameters
- PWC 150 / 170; HR, ST, ∠ST plots; Vo₂ / Vo₂max / DP; ST hysteresis; SThist, ST/HR; dQT, dPQ, dQRS
- HRV analysis on monitor segments recorded during exercise
- Athlete express assessment after Dushanin; individual card; team review
- Print strips and reports on laser / inkjet printers
- Export (Word, Excel, Matlab, raw records); e-mail
- Integration with medical information systems (HIS / MIS)
- Our own solution for telemedicine consultations, storage and data exchange — CLORM
- Interface, reports and printouts in Ukrainian, English and Russian
Documents and guides
Quick guides
Minimizing skin–electrode impedance
ECG · EEG · Holter · rheography · quick guide
1Why impedance must be low
Most of the impedance comes from the stratum corneum (dead cells, oil, dust). The higher it is, the more mains noise and baseline drift, especially on microvolt-level signals (EEG).
2Skin preparation: 5 steps
- 1Siteaway from muscle, folds, scars and wounds
- 2Hairshave or part it
- 3Cleanalcohol, let it dry completely
- 4Abraderub lightly until slightly pink
- 5Electrodefresh, with gel (RVG: moistened); press firmly, check impedance
3By examination type
ECG
- Ag/AgCl electrodes, not dried out, within expiry date
- Cable without strain; patient relaxed and not talking
EEG
- Each electrode ≤ 5 kΩ and roughly equal
- Part the hair, gel/paste, cap fitted snugly
- Check impedances in the software before recording
Holter
- Shave, always degrease and lightly abrade the skin
- Secure with tape, leave a cable loop (strain relief)
- For long recordings replace electrodes per manufacturer’s guidance
Rheography (REG, RVG)
- REG: gel electrodes over the whole contact area
- RVG: moisten the electrode/pad, do not let it dry out
- Firm fixing (strap/tape), no folds or air
✔ Pre-recording checklist
General earthing requirements for functional diagnostics systems
Earthing of functional diagnostics devices
Standards: IEC 60601-1, IEC 62353, local wiring rules
1Mains: TN-S or IT only
The protective conductor is the third (earth) pin of the socket. Never fit a PE→N jumper in the room. IT — via a medical isolating transformer.
2“Star” layout, not “daisy chain”
Each device goes to the bar with its own conductor. Loops pick up mains noise into the microvolt-level signal.
3Equipotential bonding in the room
4Earthing loop requirements
- A separate working (functional) earth, FE, independent of the hospital protective earth (protective earth norm: up to 4 Ω).
- Distance to other earthing systems — at least 15 m.
- Connect only through the third (earth) conductor of the socket.
5Testing and operation
- Measure contact resistances and insulation before first use.
- Then periodically per clinic policy (usually once a year).
- Record the results in a log.